Item: |
Universal climatic test box |
Catalogue No.: |
11566 |
WWW: |
https://strojesvoboda.com/e11566 |
Model, parameters: |
VUK 08/1500 |
Producer: |
Heraeus VÖTSCH |
Production Year: | |
Description: |
- Performance range for temperature tests:
- Temperature range:
- -80 ° C to + 180 ° C
- Temperature deviation:
- See table below
- Heating of the medium without test samples:
- 1.5 ° C / min to -70 ° C
- 1.0 ° C / min to -80 ° C
Compensation heat:
- + 20 ° C of about 1,600 W
- ± 0 ° C of about 1100W
- -20 ° C of about 800W
- -40 ° C of about 600W
- -60 ° C, 300 W
- -80 ° C 50W ca.
power range for climatic tests:
- Temperature range + 10 ° C to + 90 ° C
- Wet area: see following chart
- Wet variations: see table below
- Air: 0.1 to 4 m3 / h
Dimensions and weight:
- Dimensions in mm:
- The test area:
- A width = 1000
- Depth b = 1500
- Height c = 1000
- The weight of about 1,800 kg net
design:
- The test area:
- Stainless steel, material 1.4301
- Floor load max. 300 kg / m2
- Point loads up to 30 kg.
- Compensate for over- and underpressure
- Air guidance:
- Fan with external motor
- Sucking back
- Air on all sides
- Door:
- Opening right
- Colour: RAL 3016 (Coral)
- Inspection window:
- 500x500 mm heated
- The test area:
- Bulb in the test area - 24V, 2x40W
- Case:
- Electrically galvanized steel sheet
- Color: RAL 7032 (stone gray)
- Insulation:
- Two-layer insulation with an outer and inner vapor barrier
-cooler:
- Cascade, water cooled, 1x2,2 / 2,2 kW
- Climate system:
- BE 15 with automatic water feed
- Control:
- Two-point control for temperature and humidity measurements
- Set the desired value using potentiometer
- Digital and actual value
- An electronic controller including a sensing element Pt 100 43 760 Class B to regulate temperature and humidity
- Humidity measurement psychometrically
- Light schematic chart is signaling functionality and faults
- Safety:
- Adjustable thermal fuse
- TW temperature sensor and temperature limiter TB
- Error message - potential-free contact, max. Value of 24V, 3A
- connection
- Electrical connections:
- Nominal voltage: 380V ± 5% 3 / N 50 Hz
- Nominal power: 10 kW
- Rated current: 24 A
- Connection: 35 A plug Cekon
- Security: 35 A inertial
- The degree of protection IP 20
- Processes:
- Connecting hose for cleaning and condensate 1/2 "
- Hose for humidification and dehumidification
- Cooling water connection:
- Water supply R 3/4 "
- Drainage R 34 "
- Water pressure 3-6 bar
- The necessary pressure difference between inlet and outlet> 2.5 bar
- Consumption of cooling water, depending on the operating conditions: from 300 to 1000 l / h
- Note:
- The noise level is approximately 70 dB (A), measured at a distance of 2 m
- Permissible ambient temperature of +15 ° C to +35 ° C, max. 75% RH
- The values refer to an ambient temperature of +23 ° C ± 2 ° C and cooling water temperature of +18 ° C
- Values apply, unless otherwise specified, without test samples without irradiation and without retrofitting
- Place in a dry, ventilated room
|
All offered items are our property, physically ON OUR STOCKS.